Removing the Analyzer Tray Components
Thermo Scientific
Figure 78. Sliding the Analyzer Tray Out of the Instrument
16. Set the tray on a clean work surface.
To remove the analyzer tray components from the tray
1. Turn the analyzer tray to access the bottom and disconnect the three lens wires attached
to the lenses between the quadrupoles.
Figure 79. Finding the Lens Wires between the Quadrupoles
CAUTION INSTRUMENT DAMAGE Quadrupole surface quality is critical for instrument
performance. Take extreme care in handling the quadrupole components.
2. Remove the screws on the Q3 exit endcap holding it to the tray.
3. Remove the Q3 quad assembly and Q3 exit lens assembly from the tray.
4. Detach the Q3 exit lens assembly from the Q3 quad assembly.
5. Set the Q3 quad assembly and Q3 exit lens assembly on a clean work surface.
3
Replacing Components of the Analyzer
Lens Wires
TSQ 8000 Evo Hardware Manual
Advanced Troubleshooting
Sealing Surfaces
Analyzer Tray
Tab
113