12. Disconnect the electrometer board cable and large feedthrough cables labeled Dynode
and Electron Multiplier.
Figure 56. Disconnecting the Electron Multiplier Cables
Electrometer
Board Cable
13. Use a T20 Torxhead screwdriver to remove the four screws at each corner of the electron
multiplier plate.
CAUTION - ELECTRICAL HAZARD Avoid dropping screws between chassis and side cover
or board. If you drop a screw, find it before powering the instrument on again.
1
Performing Routine Maintenance
Replacing the Electron Multiplier
Dynode
Electron
Multiplier
Electron
Multiplier Plate
TSQ 8000 Evo Hardware Manual