Source to Ion Guide Spacer

Thermo Scientific
To replace the source to ion guide spacer:
Note See the TSQ 8000 Evo Spare Parts Guide for information about ordering this
component.
WARNING - INSTRUMENT DAMAGE Thermo Fisher Scientific strongly recommends
that only trained Thermo Fisher Scientific Field Service Engineers remove and service this
component.
1. Shut down your system by following the steps in
System"
on
page
107.
2. Remove the analyzer tray from the instrument by following the instructions in
"Removing the Analyzer Tray from the TSQ 8000 Evo Instrument"
3. Disconnect the Q1 entrance lens, Q1 exit lens, Q3 entrance lens, and Q3 exit lens from
the source interface board using a flat-tipped screwdriver.
Figure 151. Locating Wires on the Source Interface Board
3
Advanced Troubleshooting
Replacing Components of the Ion Source
"Shutting Down the TSQ 8000 Evo
on
page
TSQ 8000 Evo Hardware Manual
109.
165