Thermo Scientific
Figure 111. Replacing the Collision Cell and Quadrupole 3 in the Analyzer Tray
Q3 Quad
Assembly
Collision Cell
Assembly
16. Replace the analyzer tray into the instrument by following the instructions in
the Analyzer Tray in the TSQ 8000 Evo Instrument"
17. Restart your system by following the instructions in
3
Advanced Troubleshooting
Replacing Components of the Analyzer
Q3 Exit Lens
Assembly
Screws
on
page
118.
"Restarting the System"
on
TSQ 8000 Evo Hardware Manual
"Replacing
page
120.
135