Thermo Scientific
Figure 108. Removing the Collision Cell and Quadrupole 3 from the Analyzer Tray
Q3 Quad
Assembly
Collision Cell
Assembly
9. Replace the collision cell assembly if needed.
IMPORTANT It is critical that you use the correct collision cell, EvoCell, with the TSQ
8000 Evo MS. If you install an EvoCell in a TSQ 8000 (classic) instrument, you must
update the lens driver board.
Note If you do not need to replace quadrupole 1, go to
Figure 109. Removing Screws from the Quadrupole Rod Endcaps
Collision Cell
Assembly
10. Slide the Q1 quad assembly away from the Q1 entrance lens and endcap.
11. Replace Q1 if necessary.
12. Slide Q1 back into the Q1 entrance lens endcap.
Replacing Components of the Analyzer
step
14.
Screw to
Q1 Entrance Lens
Grounding
Endcap Screws to
Wire
Analyzer Tray
TSQ 8000 Evo Hardware Manual
3
Advanced Troubleshooting
Q3 Exit Lens
Assembly
Screws
Q1 Quad
Assembly
Q1 Entrance
Lens Endcap
133