Thermo Scientific TSQ 8000 Evo Hardware Manual page 83

Hide thumbs Also See for TSQ 8000 Evo:
Table of Contents

Advertisement

Thermo Scientific
12. Remove the small glass manifold cover from the top of the instrument.
Figure 50. Removing the Small Glass Manifold Cover
CAUTION INSTRUMENT DAMAGE: Many nitrile and latex gloves not certified for clean
room use contain silicone mold releasing agents that will contaminate the instrument. For
this reason, clean room gloves are strongly recommended when touching components
inside the vacuum manifold. We recommend Cardinal Health CP100 Nitrile Cleanroom
Gloves. See the TSQ 8000 Evo Spare Parts Guide for ordering information.
13. Look through the top of the instrument and use a clean T10 Torxhead screwdriver to
loosen the screw holding the filament retaining clip in place. Then rotate the clip away
from the filament.
1
Performing Routine Maintenance
Small Glass
Manifold Cover
TSQ 8000 Evo Hardware Manual
Replacing a Dual Filament
57