Thermo Scientific
20. Remove the axle bearing from the metal valve body. Replace the axle bearing, if necessary.
Figure 211. Replacing the Axle Bearing
Axle Bearing
Safety
Interlock
21. Pull the valve axle out of the side of the metal valve body. Replace the valve axle, if
necessary.
22. Remove the axle seal from the metal valve body. Replace the axle seal on the valve axle if
necessary.
Figure 212. Replacing the Axle Seal
Axle Seal
3
Advanced Troubleshooting
Replacing Components of the Vacuum Interlock
Clip
TSQ 8000 Evo Hardware Manual
205