Performing Routine Maintenance
Removing the Ion Source Cartridge for Cleaning
46
TSQ 8000 Evo Hardware Manual
Figure 40. Reattaching the Ion Source Cartridge to the Tool
Note The ion volume and lenses have keying features to assure the lens spring contacts
will align with the metal parts of the lenses.
5. Gently pull the source holder away from the ion source cartridge.
Figure 41. Removing the Source Holder
Tip When the source exchange tool is inverted, the source should stay attached. If this
does not happen, the source will not insert correctly into the instrument.