encounter resistance while removing the source. To release the source from the spring
contacts make small repeated rotations of the handle to the left and right quickly.
Figure 167. Pulling the Tool Down the Left Track
12. Once you reach the end of the track (when the back line at the end of the barrel is
uncovered by the black handle), twist the handle to the left to lock it into place.
Figure 168. Locking the Tool in Place
3
Advanced Troubleshooting
Replacing Components of the Ion Source
Note: Tool
Rotated Left
TSQ 8000 Evo Hardware Manual
177