3
Advanced Troubleshooting
Replacing the Boards in the TSQ 8000 Evo Instrument
272
TSQ 8000 Evo Hardware Manual
Figure 290. Disconnecting the Rod Driver Board 7-Pin Connector
Note Be careful not to bend the connector on the rod driver board on the pins on the RF
board.
9. Replace the rod driver board if necessary. Check the pin connections for accuracy when
you connect the board.
Figure 291. Replacing the Rod Driver Board
10. Attach the rod driver board to the new RF board.
11. Reattach the rod driver board's 40-pin and 7-pin connectors to the RF board.
12. Reattach the three screws and two standoffs.
7-Pin Connector
Thermo Scientific