Thermo Scientific
3. Open the manifold door and replace the tray alignment pins, magnet yoke, source
magnets, or manifold o-ring as needed on the back of the door,
Figure 252. Replacing the Tray Alignment Pins, Bracket, Manifold O-ring, and Source Magnets
Manifold O-ring
Magnet Yoke
4. Close the manifold door and reattach the four screws.
5. Restart your system by following the instructions in
Replacing Components of the Manifold
Tray Alignment Pins
"Restarting the System"
TSQ 8000 Evo Hardware Manual
3
Advanced Troubleshooting
Source Magnets
on
page
120.
233