Thermo Scientific TSQ 8000 Evo Hardware Manual page 325

Hide thumbs Also See for TSQ 8000 Evo:
Table of Contents

Advertisement

Thermo Scientific
15. Insert the new controller interface board by aligning the two connectors on the bottom of
the board to the connectors on the distribution board beneath it. It will drop into the slot
when it is in the right position.
Figure 322. Reinstalling the Controller Interface Board
16. Push the board down into the connectors.
17. Use a T20 Torxhead screwdriver to secure the two M4 screws on the controller interface
board support bracket to the chassis.
18. Reattach the top cover and left hand front panels and tighten the screws.
19. Restart your system by following the instructions in
Replacing the Boards in the TSQ 8000 Evo Instrument
Controller
Interface Board
Connectors
"Restarting the System"
TSQ 8000 Evo Hardware Manual
3
Advanced Troubleshooting
on
page
120.
299

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents