Thermo Scientific
10. Reconnect the lens heater block grounding strap to the lens heater block
Figure 158. Reconnecting the Lens Heater Block Grounding Strap
Lens
Heater
Block
Lens Heater
Grounding Strap
Screw
11. Reconnect the Q1 entrance lens, Q1 exit lens, Q3 entrance lens, and Q3 exit lens to the
source interface board using a flat-tipped screwdriver.
12. Once the components are correctly reattached, follow the instructions in
Analyzer Tray in the TSQ 8000 Evo Instrument"
TSQ instrument.
13. Restart your system by following the instructions in
3
Advanced Troubleshooting
Replacing Components of the Ion Source
"Replacing the
on
page 118
to replace the tray in the
"Restarting the System"
TSQ 8000 Evo Hardware Manual
on
page
120.
169