Thermo Scientific TSQ 8000 Evo Hardware Manual page 204

Hide thumbs Also See for TSQ 8000 Evo:
Table of Contents

Advertisement

3
Advanced Troubleshooting
Replacing Components of the Ion Source
178
TSQ 8000 Evo Hardware Manual
13. Pull the vacuum interlock handle down to close the interior ball valve.
Figure 169. Pulling Down the Vacuum Interlock Handle
14. Twist the end of the vacuum interlock handle and flip the swing handle up so that it rests
next to the vacuum interlock knob location. Then twist the end again to hold it in place.
15. Replace the plug in the VPI to prevent accidental venting and excessive leaks.
CAUTION INSTRUMENT DAMAGE: Do not move the main interlock handle when you
move the swing handle to the up position or you will open the interior ball valve, which
might cause you to lose vacuum and possibly damage the instrument.
16. Remove the tool from the instrument.