3
Advanced Troubleshooting
Replacing Components of the Ion Source

Insulating Spacers

156
TSQ 8000 Evo Hardware Manual
To replace the insulating spacers
Note See the TSQ 8000 Evo Spare Parts Guide for information about ordering this
component.
1. Shut down your system by following the steps in
System"
on
page
107.
2. Slide the top cover panel toward the back of the instrument and lift it off.
3. Use a T20 Torxhead screwdriver to remove the four screws around the manifold door.
Note Remove the screws completely so that they do not scratch the manifold when you
close the door.
4. Open the manifold door.
Figure 140. Opening the Manifold Door
"Shutting Down the TSQ 8000 Evo
Screws
Thermo Scientific