Performing Routine Maintenance
Removing the Ion Source Cartridge for Cleaning
30
TSQ 8000 Evo Hardware Manual
a. With your hand around the stainless steel barrel, twist the tool until the alignment
groove at the end of the barrel aligns with the left screw around the slot vacated by
the vacuum interlock knob.
Figure 17. Inserting the Ion Volume Tool
Alignment
Groove
b. Twist and push the stainless steel barrel to the right until it clicks into place under the
screw at the top of the slot.
Figure 18. Twisting the Tool Into Place
Alignment
Groove
Thermo Scientific