Advanced Troubleshooting
Replacing the Collision Gas Components
352
TSQ 8000 Evo Hardware Manual
21. Pull the tubing out of the manifold. Replace if necessary.
22. Disconnect the collision gas module cable from the distribution board.
Figure 378. Disconnecting the Collision Gas Module Cable from the Distribution Board
23. Disconnect the 1/16 in. tubing from the bulkhead compression fitting located on the left
hand sub panel.
Figure 379. Disconnecting the Cable to the Bulkhead Fitting
Bulkhead
Fitting
24. Use a T20 Torxhead screwdriver to remove the two screws on the collision gas module.
25. Remove the collision gas module and replace if necessary.
Collision Gas Module
Cable Connection
Collision Gas Module
Tubing
Thermo Scientific

Hide quick links: