Advanced Troubleshooting
Replacing Components of the Vacuum Interlock
200
TSQ 8000 Evo Hardware Manual
4. Disconnect the vacuum interlock cable from the connector.
Figure 202. Disconnecting the Vacuum Interlock Cable
5. Unscrew and then lower the vacuum interlock handle.
6. Pinch together the two springs on the left side the vacuum interlock shield and remove it.
Figure 203. Removing the Vacuum Interlock Shield
Springs
Lower the Vacuum Interlock Handle
Vacuum
Interlock
Cable
Connector
Vacuum
Interlock
Handle