Electrometer Board Cable - Thermo Scientific TSQ 8000 Evo Hardware Manual

Hide thumbs Also See for TSQ 8000 Evo:
Table of Contents

Advertisement

3
Advanced Troubleshooting
Replacing the Boards in the TSQ 8000 Evo Instrument

Electrometer Board Cable

316
TSQ 8000 Evo Hardware Manual
To replace the electrometer board (15-pin male/female RS-232) cable
Note See the TSQ 8000 Evo Spare Parts Guide for information about ordering this
component.
1. Shut down your system by following the steps in
System"
on
page
107. You do not have to remove the ion source cartridge or the column
or vent the instrument. If you are using hydrogen as a carrier gas, power off the GC.
2. Slide the top cover panel toward the back of the instrument and lift it off.
3. Disconnect the electrometer board cable
Figure 341. Disconnecting the Electrometer Board Cable
Electrometer
Board Cable
4. Remove all power connections from the TSQ 8000 Evo instrument.
"Shutting Down the TSQ 8000 Evo
Thermo Scientific