Left Rear Wire Mesh Filter - Thermo Scientific TSQ 8000 Evo Hardware Manual

Hide thumbs Also See for TSQ 8000 Evo:
Table of Contents

Advertisement

Left Rear Wire Mesh Filter

Thermo Scientific
To replace the left rear wire mesh filter
Note See the TSQ 8000 Evo Spare Parts Guide for ordering information.
1. Shut down your system by following the steps in
System"
on
page
107.
Note You do not have to remove the ion source cartridge or column or vent the
instrument. If you are using hydrogen as a carrier gas, power off the GC.
2. Use a T10 Torxhead screwdriver to remove the four screws holding the left rear wire mesh
filter to the left hand sub panel.
Figure 382. Replacing the Left Rear Wire Mesh Filter
3. Remove and replace the left rear wire mesh filter.
4. Reattach the four screws holding the left rear wire mesh filter to the left hand sub panel.
5. Restart your system by following the instructions in
"Shutting Down the TSQ 8000 Evo
Screws
"Restarting the System"
3
Advanced Troubleshooting
Replacing Fans and Filters
on
page
TSQ 8000 Evo Hardware Manual
120.
357