Top Manifold Cover

Thermo Scientific
To replace the top manifold cover
Note See the TSQ 8000 Evo Spare Parts Guide for information about ordering this
component.
1. Shut down your system by following the steps in
System"
on
page
107. You do not have to remove the ion source cartridge or column.
However, you must vent the instrument. If you are using hydrogen as a carrier gas, power
off the GC.
2. Slide the top cover panel toward the back of the instrument and lift it off.
3. Remove and replace the large or small glass manifold covers as needed on the top of the
instrument.
Figure 276. Replacing the Small and Large Glass Manifold Covers
4. Replace the top cover panel and tighten the screw holding it in place.
5. Restart your system by following the instructions in
Small Glass Manifold Cover
3
Advanced Troubleshooting
Replacing the Covers of the TSQ 8000 Evo Instrument
"Shutting Down the TSQ 8000 Evo
Large Glass Manifold Cover
"Restarting the System"
TSQ 8000 Evo Hardware Manual
on
page
120.
259