Thermo Scientific
21. Hold the spring and safety interlock in place as you attach the vacuum interlock shield
and screws.
Figure 246. Reattaching the Vacuum Interlock Shield
22. Reconnect the vacuum interlock cable and tubing
Figure 247. Reattaching the Vacuum Interlock Cable and Vacuum Interlock Knob
3
Advanced Troubleshooting
Replacing Components of the Manifold
Screws
Vacuum Interlock
Shield
Vacuum
Interlock
Knob
Vacuum
Interlock
Cable
Connector
TSQ 8000 Evo Hardware Manual
227