5. Use a T10 Torxhead screwdriver to remove the four screws holding the electrometer
board in place.
Figure 267. Removing the Electrometer Board
Screws
6. Remove the electrometer board to reveal the anode feedthrough.
Figure 268. Removing the Anode Feedthrough, O-ring, Standoff and Screws
Anode Feedthrough
and O-ring
Screws