Upgrade Equipment
CI Reagent Gas Flow Module
392
TSQ 8000 Evo Hardware Manual
27. Connect the Ethernet cable to the PC communication board.
28. Secure the left hand sub panel to the chassis with the three screws.
29. Plug in the power connections.
30. Plumb the reagent gas from your gas supply to either port on your CI gas controller.
WARNING FIRE HAZARD: Some CI gases, such as methane and isobutane, are
flammable. Make sure these gases are properly exhausted and all gas fittings on the system
are leak-free. Consult your local Environmental and Safety Regulations for information
about how to properly exhaust fumes from your laboratory.
WARNING TOXIC SUBSTANCES HAZARD: Some CI gases, such as ammonia, are toxic.
Make sure these gases are properly exhausted and all gas fittings on the system are
leak-free. Consult your local Environmental and Safety Regulations for information about
how to properly exhaust fumes from your laboratory.
CAUTION INSTRUMENT DAMAGE: Do not exceed 240 kPa (35 psig) or you could
damage the CI reagent gas flow module.
31. Turn on the reagent gas at the gas supply.
32. Check for leaks at each reagent gas fitting.
33. Reattach the left hand front panel and top cover panel and tighten the screws.
34. Reattach the top cover panel and left side panel and tighten the screws.
35. Restart your system by following the instructions in
Note Purge air fro the CI reagent gas supply tubing overnight before using your TSQ
8000 instrument to acquire data.
"Restarting the System"
on
Thermo Scientific
page
120.