Thermo Scientific TSQ 8000 Evo Hardware Manual page 190

Hide thumbs Also See for TSQ 8000 Evo:
Table of Contents

Advertisement

3
Advanced Troubleshooting
Replacing Components of the Ion Source
164
TSQ 8000 Evo Hardware Manual
Figure 150. Locating Wires on the Source Interface Board
5. Attach a new lens plate and springs to the lens heater block and reattach the screws.
6. Reattach the lens 1, lens 2, and lens 3 wires from the source interface board using a
flat-tipped screwdriver.
7. Once the components are correctly reattached, follow the instructions in
Analyzer Tray in the TSQ 8000 Evo Instrument"
TSQ instrument.
8. Restart your system by following the instructions in
"Replacing the
on
page 118
to replace the tray in the
"Restarting the System"
on
page
120.