Thermo Scientific TSQ 8000 Evo Hardware Manual page 241

Hide thumbs Also See for TSQ 8000 Evo:
Table of Contents

Advertisement

Thermo Scientific
5. Look through the left side of the instrument and use a T20 Torxhead screwdriver to
loosen the screw holding the board to the chassis.
Tip The bracket has a white diagram describing the location of this screw. It is
approximately two inches from the top of the board directly under the white label.
Figure 225. Removing the Lens Driver Board
6. Use your fingers to carefully pull the bracket of the board away from the chassis and
toward the left side of the instrument, until the 20 pin connector clears the feedthrough.
7. Pinch together the two springs on the left side the vacuum interlock shield to free the
hinge from the vacuum interlock shield.
Figure 226. Freeing the Hinge from the Vacuum Interlock Shield
Springs
3
Advanced Troubleshooting
Replacing Components of the Manifold
Lens Driver Board
Screw
TSQ 8000 Evo Hardware Manual
215