Thermo Scientific TSQ 8000 Evo Hardware Manual page 149

Hide thumbs Also See for TSQ 8000 Evo:
Table of Contents

Advertisement

Thermo Scientific
11. Look inside the endcap and use a T10 Torxhead screwdriver to remove the two screws
holding the ion guide to the Q1 entrance lens endcap and Q1 entrance lens.
Figure 93. Finding the Quad Entrance Lens
Screws
12. Rotate the ion clamp until the ion guide can be separated from the endcap
13. Pull the ion guide clamp away from the ion guide.
14. Place the ion guide clamp over the new ion guide.
Figure 94. Replacing the Ion Guide
15. Insert the new ion guide into the Q1 entrance lens endcap.
16. Rotate the clamp until the ion guide is locked in place. Make sure the ion guide is aligned
parallel to the flat on the endcap.
17. Tighten the two T10 screws inside the endcap.
Note Make sure the Q1 entrance lens sits flush and centered in the Q1 entrance lens
endcap.
Ion Guide Wire
3
Advanced Troubleshooting
Replacing Components of the Analyzer
Q1 Entrance
Lens Endcap
Q1 Entrance Lens
TSQ 8000 Evo Hardware Manual
123

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents