Investigating Power Supply Issues

TSQ 8000 Evo system does not power on.

Table 24. Possible Remedies when TSQ 8000 Evo System does not Power On
Make sure the TSQ 8000 Evo instrument's power cable is properly connected to the instrument and electrical outlet.
Verify that the electrical outlet is functioning properly.
Replace the TSQ 8000 Evo instrument's power supply system. See
TSQ 8000 Evo system powers on temporarily and trips the circuit breaker.
Table 25. Possible Remedies when TSQ 8000 Evo System Powers On Temporarily and Trips the Circuit Breaker
Make sure the instrument's power cable is not damaged.
Check the foreline pump voltage setting and replace the pump, if necessary.
Replace the TSQ 8000 Evo instrument's power supply system. See
±RF DC supply is outside the acceptable range.
Table 26. Possible Remedy when the ±RF DC Supply is Outside the Acceptable Range
Make sure the Rod Driver board is properly connected and use Xcalibur Power Status to ensure the readback reports +48 V DC is present.
Thermo Scientific
When you run Power Supply Check diagnostics on the TSQ 8000 Evo mass spectrometer, the
Xcalibur software verifies that all power supply voltages are within the acceptable ranges. Your
instrument or software may indicate the following types of issues.
WARNING
ELECTRICAL SHOCK HAZARD: When troubleshooting any issue that
-
requires removing a cover on the TSQ 8000 Evo mass spectrometer, you should power-off
and vent the instrument to avoid harming yourself.