Thermo Scientific
Figure 388. Reinstalling the Lens Driver Board
Connector for
Lens Driver
Board
21. Look through the left side of the instrument and use a T20 Torxhead screwdriver to
tighten the screw holding the lens driver board in place on the chassis.
22. Reattach the left hand front panel and tighten the four screws holding it in place.
23. Reattach the top cover panel and tighten the screw holding it in place.
24. Restart your system by following the instructions in
3
Advanced Troubleshooting
Replacing Fans and Filters
"Restarting the System"
on
TSQ 8000 Evo Hardware Manual
page
120.
363