Thermo Scientific
Figure 82. Removing the Analyzer Tray Screws
Ion Guide Screws
10. Remove the Q1 entrance lens wire from the source interface board.
11. Remove the ion guide assembly screws.
12. Remove the ion guide assembly.
13. Use a T10 Torxhead screwdriver to remove the lens heater block grounding strap from the
analyzer tray.
14. Remove the Q1 exit, Q3 entrance, and Q1 exit lens wires from the analyzer tray. See
Figure
79.
15. Use a T20 Torxhead screwdriver to remove the two screws attaching the source block to
the analyzer tray.
IMPORTANT Be careful not to drop the MACOR ceramic plate. See
fragile.
Figure 83. Removing the Ion Source from the Analyzer Tray
Screws
16. Remove the source block from the analyzer tray and set it aside.
Replacing Components of the Analyzer
Q3 Entrance Endcap Screw
Q1 Exit Endcap
Screw
TSQ 8000 Evo Hardware Manual
3
Advanced Troubleshooting
Q3 Exit Endcap
Screw
Figure
83. It is
MACOR
Ceramic Plate
115