4. Inject the FC 43. If you see liquid pooled on top of the white frit, remove the excess
liquid according to local environmental regulations.
CAUTION INSTRUMENT DAMAGE: Adding more than 300 μL of calibration compound
can damage the calibration gas controller. Be sure liquid does not get into the controller
when you reattach the reservoir.
5. Reattach the cover to the calibration gas controller.
6. Close the front door of the instrument.
1
Performing Routine Maintenance
Maintaining the Calibration Gas Module
TSQ 8000 Evo Hardware Manual