11. Twist the vent valve knob to remove it and the vent valve o-ring.
Figure 236. Removing the Vent Valve Knob and O-Ring
Vent Valve
O-ring
Vent
Valve
Knob
12. Flip the manifold door over and remove the manifold o-ring.
Figure 237. Removing the Manifold O-Ring
Manifold
O-ring
13. Twist and remove the tray alignment pins that hold the magnet yoke to the back of the
manifold door. The magnet yoke will pull free of the door, releasing the alignment collar.
Figure 238. Removing the Tray Alignment Pins and Bracket
Magnet Yoke
Tray Alignment
Pins
3
Advanced Troubleshooting
Replacing Components of the Manifold
TSQ 8000 Evo Hardware Manual
223