Thermo Scientific TSQ 8000 Evo Hardware Manual page 96

Hide thumbs Also See for TSQ 8000 Evo:
Table of Contents

Advertisement

1
Performing Routine Maintenance
Cleaning Durable Components
70
TSQ 8000 Evo Hardware Manual
Tip Clean the CI ion volume under a microscope to verify that you removed all
debris.
f.
If you notice any scratches on the components, you may need to replace the
component. Scratches can affect the performance of your instrument.
2. Rinse the components with clean tap water. Use a toothbrush under a stream of water to
remove the aluminum oxide slurry. Do not let the slurry dry on the components because
it is difficult to remove. If the components still look dirty, repeat step 1.
3. Sonicate the components in warm detergent.
a. Use forceps to place the components in a beaker of warm detergent.
b. Place the beaker and its contents in an ultrasonic bath for five minutes.
c. Rinse the components with tap water to remove the detergent.
4. Sonicate the components in deionized water.
a. Use forceps to place the components in a beaker of deionized water.
b. Place the beaker and its contents in an ultrasonic bath for five minutes. If water is
cloudy after sonicating, replace the water with fresh water, and put the beaker and
contents in a ultrasonic bath again for five minutes. Repeat until the water is clear.
5. Use forceps to immediately transfer the components to a clean beaker of acetone.