Advanced Troubleshooting
Replacing the Boards in the TSQ 8000 Evo Instrument
310
TSQ 8000 Evo Hardware Manual
Figure 334. Disconnecting the Cable to the Bulkhead Fitting
Bulkhead
Fitting
28. Disconnect the collision gas module cable from the distribution board.
29. If you are using CI, follow the instructions in the
remove the CI reagent gas flow module.
30. Now that all of the components have been removed from the distribution board, you can
see that it lies flat on the bottom of the chassis.
Collision Gas Module
Tubing
CI Reagent Gas Flow Module
section to