Thermo Scientific
6. Insert the ion volume with the handles fitting into the notches of the sleeve. Make sure
the ion volume is firmly seated into the gap on lens 1. You may need to rotate lens 1
slightly to make the ion volume fit correctly.
Figure 180. Inserting the Ion Volume into the Source Sleeve
Note The ion volume handles are different sizes and will only fit into the sleeve one way.
7. Insert the ion volume repeller insulator onto the repeller.
8. Tighten the repeller nut.
Replacing Components of the Ion Source
Handles Fit
Here
TSQ 8000 Evo Hardware Manual
3
Advanced Troubleshooting
Ion Volume
Handles
Lens 1 Gap
185