7. Move the TSQ 8000 Evo instrument away from the GC so you can access the transfer
line on the right side.
WARNING BURN HAZARD: The transfer line may be hot, so be careful.
8. Use a T20 Torxhead screwdriver to remove the five screws on the right side panel.
Figure 371. Removing the Right Side Panel Screws
9. Slide the right side panel to the right of the instrument and lift it off.
10. Remove the rear RF board. The rear RF board is the third board standing on end toward
the center of the TSQ 8000 Evo instrument. Use a T20 Torxhead screwdriver to remove
the screw holding the board and bracket to the chassis.
CAUTION - INSTRUMENT DAMAGE Do not touch the components of the boards. Hold
the boards by the sheet metal bracket.
Replacing the Collision Gas Components
Right Side
Panel Screws
TSQ 8000 Evo Hardware Manual
3
Advanced Troubleshooting
347