Thermo Scientific
12. Use a T20 Torxhead screwdriver to remove the two screws on the back of the source
heater block.
Figure 123. Disconnecting the Ion Source to the Analyzer Tray
Screws
Note Some of the pieces may become slack as you loosen the screws. Do not let them fall.
13. Slide the source interface board and the source block out from under the tray without
damaging or disconnecting the wires.
CAUTION - INSTRUMENT DAMAGE The lens/source heater may be fused to the source
block and difficult to remove. To avoid damage, separate the block from the tray before
removing the lens/source heater.
Figure 124. Removing the Ion Source from the Analyzer Tray
Wires
3
Advanced Troubleshooting
Replacing Components of the Ion Source
Ion Source
Block
TSQ 8000 Evo Hardware Manual
145