TSQ 8000 Evo Hardware Manual
11. Use a T20 Torxhead screwdriver to remove the two screws holding the support bracket to
the bottom of the instrument.
Figure 413. Removing the Controller Interface Board Support Bracket
12. Disconnect the Ethernet cable from the PC communication board.
13. Remove the controller interface board and PC communication board from the
instrument. The two boards are connected. Do not separate them.
Support
Bracket
Screws