Thermo Scientific TSQ 8000 Evo Hardware Manual page 266

Hide thumbs Also See for TSQ 8000 Evo:
Table of Contents

Advertisement

3
Advanced Troubleshooting
Replacing Components of the Electron Multiplier
240
TSQ 8000 Evo Hardware Manual
6. Use a T10 Torxhead screwdriver to remove the anode feedthrough, two screws, standoff.
and o-ring. Set them aside on a clean work surface.
Figure 260. Removing the Anode Feedthrough
Anode
Feedthrough
and O-ring
7. Use a T20 Torxhead screwdriver to remove the four screws at each corner of the electron
multiplier plate.
Figure 261. Disconnecting the Electron Multiplier Plate
Electron Multiplier
Plate Screws
Standoff
Screws
Thermo Scientific