9. Replace the manifold door hinge and reattach the screws.
10. From the front of the instrument, reattach the hinge spring to the vacuum interlock
shield.
11. Align the connector on the bottom of the lens driver board with the connectors on the
distribution board beneath it. It will click into place when its in the right position.
Figure 228. Reinstalling the Lens Driver Board
Connector
3
Advanced Troubleshooting
Replacing Components of the Manifold
TSQ 8000 Evo Hardware Manual
217