Advanced Troubleshooting
Replacing the Calibration Gas Components
340
TSQ 8000 Evo Hardware Manual
5. Attach a new calibrant reservoir to the calibration gas module.
Figure 363. Replacing the Calibrant Reservoir and Cover
Calibrant
Reservoir
6. Fill the calibrant reservoir of the new module by following the instructions in
the Calibrant Reservoir"
7. Reattach the calibrant reservoir cover.
8. Close the front door of the instrument.
9. If you are using hydrogen as a carrier gas, attach the hydrogen safety screws and restart
your GC by following the instructions in
on
page
52.