Malfunctions; Error Messages - Sirona Sinius CS Operating Instructions Manual

Treatment center
Hide thumbs Also See for Sinius CS:
Table of Contents

Advertisement

Sirona Dental Systems GmbH
Operating Instructions SINIUS / SINIUS CS
Code
Error/fault
10
Flow rate of SPRAYVIT M
on dentist element too low
11
Flow rate of SPRAYVIT M
on assistant element too
low
12
Tumbler temperature fuse
defective
13
Battery of wireless foot
control empty
14
Disinfectant consumption
too low
15
Water feed too low
17
Emergency pump operation Malfunctioning of pump sensors in the
18
Chair movement outside of
the permissible movement
limits or chair drive position
not plausible
19
Injection valve for
disinfectant defective
63 22 726 D3561
D3561.201.01.03.02 05.2012
7

Malfunctions

Malfunctions
7.1

Error messages

Any error states of the treatment center that cannot be immediately
Error messages
recognized by the user and must nevertheless be corrected are displayed
on the touchscreen. If an error state exists, this is indicated by a warning
triangle (A) in the status line of the touchscreen.
➢ Touch the warning triangle (A) in the footer of the touchscreen.
 The error code(s) is(/are) displayed.
The error codes have the following meaning:
Error table, without HF surgery, without hydrocolloid, without safety switch for chair frame
Description
The water flow rate is below the limit
value. The heater of the SPRAYVIT M
is no longer heated to prevent
scalding by overheated water.
The temperature fuse was blown by
overheating. The water in the tumbler
filling unit is no longer heated.
Operation of the treatment center no
longer possible due to fully depleted
battery.
The disinfectant tank was not emptied
since its last filling even though many
disinfection cycles were performed.
The maximum filling time for the
mixing tank is exceeded.
water unit.
If the chair is located outside of the
permissible movement limits or the
position is not plausible, the chair
stops moving.
No disinfectant is added if the
injection valve is defective.
7.1 Error messages
Action
Clean the outlet nozzle of the
SPRAYVIT, see the "SPRAYVIT M
Operating Instructions" and perform
the flow test, see "Checking the flow
rate on the SPRAYVIT M
multifunctional syringe".
Inform your service engineer.
See "Changing the battery of the
wireless foot control" [ → 215].
Inform your service engineer.
Change the water filter, see
"Changing the water and air
filters" [ → 201].
Check the water pressure of the
drinking water network.
Inform your service engineer.
Inform your service engineer.
Inform your service engineer.
7 Malfunctions