6 Disinfection system
6.3 Sanitation
6.3
Sanitation
Sanitation intervals
Special measures for additional devices
20
A
DESINF
Care and cleaning instructions for the practice team C2
Perform sanitation every 4 weeks or whenever the
germination index significantly exceeds 100 germs
per ml (see section 6.1 "Microbiological water test").
This is the only way to combat the formation of bio-
film in the water paths safely and effectively.
Special devices that are connected to the chair via the
media block are treated with disinfected water with a
higher hydrogen peroxide concentration (1.4%) during
sanitation.
ATTENTION
If the connected devices are not suitable for this higher
sanitation concentration, Sirona recommends discon-
necting them from the treatment center during sanitation
and performing a separate sanitation in accordance with
the manufacturer's specifications.
Preparations
Preparation for sanitation takes approx. 10 minutes.
•
If DESINF flashes on the dentist element display,
first add 1 liter of DENTOSEPT P (see the Operating
Instructionsfor the , section on "Disinfection system
(optional)").
The flashing DESINF message then disappears.
For reorders:
DENTOSEPT P
For Canada/Japan
only:
•
Maximum water flow must be set on all burr drives
as well as on the SIROSONIC L. Check this before
every sanitation procedure.
•
All instruments and suction hoses must then be
placed in their holders.
Sirona Dental Systems GmbH
+
+
, M1
1 case = 6x1 liter,
Order No. 33 18 156
1 case = 6 x 1 liter,
Order No. 59 69 535
59 58 512 D 3370
D 3370.103.01.13.02