Sirona Dental Systems GmbH
Installation Requirements SINIUS / SINIUS CS / SINIUS TS
Unevenness
Load capacity:
A
63 22 668 D3561
D3561.021.01.06.02 12.2014
3
On-site installation
3.1
Substrate, floor
The floor must be level and horizontal in accordance with DIN 18 202.
Level floor, steel adapter plate
In the case of an uneven floor, the steel adapter plate must be used; see
Mounting plates [ → 38].
3.1.1
Load-bearing capacity of the floor
The minimum load-bearing capacity of the floor must be 0.5 N/cm
(corresponds to around 500 kg/m
0,5
0,7
2
N/cm
lbs/sq in.
3.1.2
Tensile forces
The floor must be able to accommodate tensile forces of > 1 kN at each
of the marked screw fastening points (A).
It may be necessary to reinforce the surface beneath the floor.
When using a mounting plate (see Mounting plates [ → 38]) this
requirement is mandatory.