1
Installing the GC
Install hydrogen sensor calibration gas
If available, install the hydrogen sensor calibration gas. The hydrogen sensor ship kit includes a
pressure regulator that installs directly onto the calibration gas cylinder. Agilent also provides
sufficient copper tubing and hardware to connect the pressure regulator to the hydrogen
sensor assembly calibration gas input fitting.
1 Set the gas cylinder stand upright on the lab bench near the GC.
The gas cylinder is pressurized. Do not place the cylinder on its side or behind the oven
WAR N IN G
exhaust vent!
Pressure
regulator
Calibration
gas cylinder
Gas cylinder
stand
2 Turn the pressure regulator completely off and set the output pressure as low as possible
3 Install the pressure regulator onto the calibration gas cylinder (it screws on).
34
(full CCW).
1/8-inch Swagelock
nut and ferrule
To hydrogen sensor
input fitting
1/8-inch copper
tubing, cut to
length needed
Installation and First Startup