Figure 82. Diagnostics Dialog Box Showing The Set Device List - Thermo Scientific LTQ Series Hardware Manual

Table of Contents

Advertisement

Thermo Scientific
3. (Velos Pro only) Place a multiplier shield retainer on the base of each multiplier as shown
in
Figure
81.
• For the left multiplier, place the shield retainer over the left-side screw holes.
• For the right multiplier, place the shield retainer over the right-side screw holes.
4. Using the 5/64 in. hex driver, secure each electron multiplier assembly to the top cover
plate with the hex socket-head screws and flat washers, if provided. Use the screw hole
location specified in
Ensure that the shield retainers remain tightly positioned against the multipliers' bases.
5. Remove the tissue that is over the vacuum manifold opening.
6. Reinstall the top cover plate over the vacuum manifold opening (see
7. Reinstall the top cover of the mass spectrometer.
8. Start up the system as described in
page
37.
After replacing the electron multipliers, reset and save the initial voltage settings to the
appropriate values for new multipliers, and then calibrate the electron multipliers' voltages.
To set the electron multiplier voltages
1. Open the Tune Plus window (see
2. Choose Diagnostics > Diagnostics to open the Diagnostics dialog box
3. In the Tools list, select Set Device.

Figure 82. Diagnostics dialog box showing the Set Device list

Figure
80.
"Starting the System after a Complete Shutdown"
page
34).
5
Maintenance
Electron Multiplier Replacement
page
115).
(Figure
82).
LTQ Series Hardware Manual
on
119