Loop Test

Thermo Fisher Scientific
Chapter 6
Diagnostics
This chapter provides on the Diag/Service menu in the NI-FBUS
Configurator root menu. The Diag/Service menu provides access to the
Loop Test, Trim Analog Output, and Alarm Bad Flags Status items.
Note If you are not using the NI-FBUS Configurator, reference the
documentation that came with your host system.
The Loop Test function allows you to manually manipulate the analog
output to a selected constant output value.
1. In the NI-FBUS Configurator, go to AI_TRANSDUCER > Root
Menu > Diag/Service > 1 Loop Test > Loop test.
2. The first loop test screen appears. Click Next when you are ready to
continue.
Figure 6–1. Initial loop test screen
LevelPRO with F
™ fieldbus Application Guide
OUNDATION
6-1