Problem
Display appears dim
Screen not displaying
correct data
No data on disk
Diagnostics
Thermo Fisher Scientific
Checkpoint
Verify the temperature and
humidity are within
specifications.
Is screen dimmed due to
period of inactivity?
Intensity level setting correct?
Is recorder in Live mode?
Is digital filter constant 0?
Is input signal's amplitude
fluctuating rapidly?
Recorder is in Live mode,
digital filter constant is 0, and
input signal's amplitude is not
fluctuating rapidly.
Is proper media type selected?
Is disk formatted correctly?
Is history function enabled
and record mode on?
Can you save configuration
file to media?
From the Program > System > Diagnostics screen, you can perform a
certain amount of diagnostics on the recorder from the front display.
Maintenance & Troubleshooting
Yes/No
Solution
Temperature: -10°C to 50°C (14°F to
122°F)
Humidity: 10 to 90% RH, non-condensing
Yes
Touch display.
No
Adjust brightness level setting (Chapter 4).
No
Adjust intensity level setting (Chapter 4).
No
Switch to Live mode (Chapter 4).
No
Check filter process.
Yes
Verify recorder can follow input.
Contact Thermo Scientific.
No
Select proper media type (Chapter 4).
No
Format disk.
No
Enable history function (Chapter 4) and
recording (Chapter 9).
Yes
Check history and recording functions.
No
Verify installed Compact Flash
Device is in accordance with Table
1-10.
Replace media drive.