Optional Analog Port/Calibration - Optional - Thermo Scientific NESLAB EX Series Installation, Operation And Maintanance Manual

Bath/circulator
Hide thumbs Also See for NESLAB EX Series:
Table of Contents

Advertisement

Optional Analog Port/
Calibration (Digital Plus Only)
P ress and Hold
No
XX.X
Yes
No
Release
No
CAL
Yes
rtd1
No
rdt3
No
A in
Yes
No
No
AoUt
Yes
No
No
No
S torE
Yes
Accept All Changes
The analog signal is setup using the unit's Computer Setup Loop. With the
bath on, press and hold the Computer button for five seconds. The controller
will display SErL. If needed, press the arrow buttons to toggle between 232,
485, and A In. With A In displayed, press the Computer button again until
StorE is displayed, press YES. The Computer button will flash when A in is
enabled. Use the Computer button to enable/disable the analog signal.
The analog input and output are calibrated at the factory. If you need to
recalibrate follow the diagram until the display indicates A in.
Pressing YES at the A in display and the display will indicate A iH. Press
YES and the display will alternate between A iH and some value. Apply 2.5V
to the analog input pin 15, see page 27 (analog ground is pin 6). Measure the
precise value with a voltmeter.
Change the value on the bath display to match the voltmeter reading.
A iH
A iH
Yes
No
A iL
A iL
Yes
A oH
A oH
Yes
No
A oL
Yes
Company preset values:
A iH = 2.5000
A iL = 0.0000
A oH = 2.5000
A oL = -.5000
At the StorE display press YES to save the calibration, press No to abort it.
NOTE: If error code Er16 appears you will need to recalibrate.
Press SCROLL to check your value and
Yes
x.xxxx
then press SCROLL again to display A iL.
No
If desired, repeat the procedure for A iL.
Yes
x.xxxx
Apply 0V to pin 15. Measure the voltage
No
and change the displayed value to match it
Pressing YES at the AoUt display and the
Yes
x.xxxx
display will indicate A oH. Press YES and
No
the display will alternate between A oH
and 2.5000, the bath will set the Temp Out
pin 7 output voltage to approximately 2.5V.
Yes
x.xxxx
Measure the voltage and change the
No
displayed value to match it. Press
SCROLL to check your value and then
press SCROLL again to display A oL.
If desired, repeat the procedure for A oL.
The display will alternate between A oH
and -.5000, the bath will set the Temp Out
pin 7 output voltage to approximately
-0.5V. Measure the voltage and change the
value to match the meter.
- 38 -