:SAVE:WAVeform:LENGth

Command Syntax
:SAVE:WAVeform:LENGth <length>
<length> ::= 100 to max. length; an integer in NR1 format
When the :SAVE:WAVeform:LENGth:MAX setting is OFF, the
:SAVE:WAVeform:LENGth command sets the waveform data length (that is,
the number of points saved).
When the :SAVE:WAVeform:LENGth:MAX setting is ON, the
:SAVE:WAVeform:LENGth setting has no effect.
Query Syntax
:SAVE:WAVeform:LENGth?
The :SAVE:WAVeform:LENGth? query returns the current waveform data
length setting.
Return Format
<length><NL>
<length> ::= 100 to max. length; an integer in NR1 format
See Also
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
(see
page
1138)
"Introduction to :SAVE Commands"
":SAVE:WAVeform:LENGth:MAX"
":SAVE:WAVeform[:STARt]"
":WAVeform:POINts"
on page 972
":SAVE:WAVeform:FORMat"
on page 607
on page 624
on page 621
on page 622
26
:SAVE Commands